High Accuracy, High Performance and Reliable Measurements

  •  0.02% basic DC accuracy
  • Large display with 50,000 counts
  • Dual display for double indication for AC and DC, V and Hz, etc.
  • True-RMS Measurements AC and AC+DC
  • Wide AC Frequency bandwidth from 10Hz to 100kHz
  • True-RMS or MEAN value detection mode can be selected
  • Fast Peak Hold response time of 250µs
  • Low-pass filter for motor drive measurements
  • Low Power Ω for resistance measurements on sensible electronic circuits by a low and safe test current
  • User calibration function

Safety design for industrial use

  • Complies with IEC 61010-1 CAT IV 600V, CAT III 1000V
  • Terminal shutter to prevent incorrect test leads' insertion in current terminals
  • Very wide operating temperature range from -20 to 55ºC

Reliable support for data management

  • Large data internal memory
    KEW 1062: 10,000 data (logging), 100 data (manual)
  • Download data and Live Monitoring on a PC via the USB interface
    (Option for USB Communication set)
    * MAX 32,767 data can be transferred.

Specification

Detection mode MEAN/Ture RMS (switch)
DC V 50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V
(Input impedance: Approx. 100MΩ [50/500/2400mV], 10MΩ [5/50/500/1000V])
±0.02%rdg±2dgt (Basic accuracy)
AC V [Ture RMS] 50.000/500.00mV/5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V])
±0.4%rdg±30dgt (Basic accuracy)
AC V [MEAN] 50.000/500.00mV/ 5.0000/50.000/500.00/1000.0V
(Input impedance:11MΩ<50pF [50/500mV/5V],10MΩ<50pF [50/500/1000V])
±1%rdg±30dgt (Basic accuracy)
DCV+ACV 5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [5V], 10MΩ<50pF [50/500/1000V])
±0.5%rdg±10dgt (Basic accuracy)
DC A 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
±0.2%rdg±5dgt (Basic accuracy)
AC A [Ture RMS] 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
±0.75%rdg±20dgt (Basic accuracy)
AC A [MEAN] 500.00/5000.0µA/50.000/500.00mA/ 5.0000/10.000A
±1.5%rdg±20dgt (Basic accuracy)
DCA+ACA 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
±1%rdg±10dgt (Basic accuracy)
Ω 500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ
±0.05%rdg±2dgt (Basic accuracy)
LowPower-Ω 5.000/50.00/500.0kΩ/5.000MΩ
±0.2%rdg±3dgt (Basic accuracy)
Continuity buzzer 500.0Ω
(The buzzer turns on at resistances lower than 100±50Ω)
Diode test 2.4000V
±1%rdg±2dgt
Open curcuit voltage: <5V (Approx. 0.5mA Measuring Current)
Capacitance 5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF
±1%rdg±5dgt (Basic accuracy)
Frequency 2.000 - 9.999/9.00 - 99.99/90.0 - 999.9Hz/0.900 - 9.999/9.00 - 99.99kHz
±0.02%rdg±1dgt (Basic accuracy)
DUTY 10 - 90%
±1%rdg
Temperature -200 - 1372ºC
±1%rdg±1.5ºC
-328 - 2501.6ºF
±1%rdg±2.0ºF (with the use of K-type Temperature probe)
Measurement functions DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-Ω
Other functions

Data Hold (D/H), Auto Hold (A/H), Peak Hold (P/H), Range Hold (R/H), Maximum value (MAX), Minimum value (MIN)
Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight

Applicable standards IEC 61010-1 CAT IV 600V / CAT III 1000V Pollution degree 2
IEC 61010-031, IEC 61326-1 (EMC)
Power source R6 (1.5V)×4
(Auto power off: approx. 20 minutes)
Dimension 192(L)×90(W)×49(D) mm
Weight Approx. 560g (including batteries)
Accessories

7220A (Test leads)
8926 (Fuse [440mA/1000V])×1 (included)
8927 (Fuse [10A/1000V])×1 (included)
R6×4
Instruction manual

Optional accessories

7234 (Alligator clip)
8241 (USB Communication set)
8115 (AC/DC Clamp Sensor)
8121 (Load current Clamp Sensor)
8122 (Load current Clamp Sensor)
8123 (Load current Clamp Sensor)
8146 (Leakage & Load current Clamp Sensor)
8147 (Leakage & Load current Clamp Sensor)
8148 (Leakage & Load current Clamp Sensor)
7146 (Banana Ø4 adjuster plug)
8405 (Temperature probe)
8406 (Temperature probe)
8407 (Temperature probe)
8408 (Temperature probe)
9154 (Carring case)